Vulnerability identifier: #VU100583
Vulnerability risk: Low
CVSSv4.0: 6.1 [CVSS:4.0/AV:L/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:U/U:Clear]
CVE-ID:
CWE-ID:
CWE-307
Exploitation vector: Local
Exploit availability: No
Vulnerable software:
Life2000 Ventilation System
Hardware solutions /
Other hardware appliances
Vendor: Baxter
Description
The vulnerability allows a local attacker to compromise the target system.
The vulnerability exists due to improper restriction of excessive authentication attempts within the Clinician Password or the Serial Number Clinician Password. A local attacker can conduct brute force attacks to gain unauthorized access to the ventilator and gain access to the system.
Mitigation
Cybersecurity Help is currently unaware of any official solution to address this vulnerability.
Vulnerable software versions
Life2000 Ventilation System: 06.08.00.00
External links
https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01
https://www.baxter.com/sites/g/files/ebysai3896/files/2024-11/ICSMA-24-319-01-Baxter-Life-2000-Ventilation-System.pdf
Can this vulnerability be exploited remotely?
No. This vulnerability can be exploited locally. The attacker should have authentication credentials and successfully authenticate on the system.
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.