#VU56995 Reliance on Component That is Not Updateable in ZOOM LATITUDE Programmer/Recorder/Monitor Model 3120


Published: 2021-10-01

Vulnerability identifier: #VU56995

Vulnerability risk: Low

CVSSv3.1: 6 [CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:C/C:L/I:H/A:L/E:U/RL:U/RC:C]

CVE-ID: CVE-2021-38398

CWE-ID: CWE-1329

Exploitation vector: Local

Exploit availability: No

Vulnerable software:
ZOOM LATITUDE Programmer/Recorder/Monitor Model 3120
Hardware solutions / Medical equipment

Vendor: Boston Scientific

Description

The vulnerability allows a local attacker to compromise the target system.

The vulnerability exists due to the affected device uses off-the-shelf software components that contain unpatched vulnerabilities. An attacker with physical access can exploit these vulnerabilities.

Mitigation
Cybersecurity Help is currently unaware of any official solution to address this vulnerability.

Vulnerable software versions

ZOOM LATITUDE Programmer/Recorder/Monitor Model 3120: All versions


External links
http://ics-cert.us-cert.gov/advisories/icsma-21-273-01


Q & A

Can this vulnerability be exploited remotely?

No. The attacker should have physical access to the system in order to successfully exploit this vulnerability.

Is there known malware, which exploits this vulnerability?

No. We are not aware of malware exploiting this vulnerability.