Risk | Medium |
Patch available | NO |
Number of vulnerabilities | 1 |
CVE-ID | CVE-2019-10966 |
CWE-ID | CWE-287 |
Exploitation vector | Network |
Public exploit | N/A |
Vulnerable software Subscribe |
Aestiva Anesthesia Hardware solutions / Medical equipment Aespire Anesthesia Hardware solutions / Medical equipment |
Vendor | GE Healthcare |
Security Bulletin
This security bulletin contains one medium risk vulnerability.
EUVDB-ID: #VU19170
Risk: Medium
CVSSv3.1: 4.9 [CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:N/E:U/RL:U/RC:C]
CVE-ID: CVE-2019-10966
CWE-ID:
CWE-287 - Improper Authentication
Exploit availability: No
DescriptionThe vulnerability allows a remote attacker to bypass authentication process.
The vulnerability exists due to an error when serial devices are connected via an added unsecured terminal server to a TCP/IP network configuration. A remote attacker can gain unauthorized access to the application, modify device donfiguration and silence alarms.
MitigationCybersecurity Help is currently unaware of any official solution to address this vulnerability.
Aestiva Anesthesia: 7100 - 7900
Aespire Anesthesia: 7100 - 7900
External linkshttp://www.us-cert.gov/ics/advisories/icsma-19-190-01
Q & A
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the Internet.
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.