Exposure of resource to wrong sphere in Philips IntelliSpace Perinatal



Published: 2019-10-25
Risk Low
Patch available NO
Number of vulnerabilities 1
CVE-ID CVE-2019-13546
CWE-ID N/A
Exploitation vector Local
Public exploit N/A
Vulnerable software
Subscribe
IntelliSpace Perinatal
Hardware solutions / Medical equipment

Vendor Philips

Security Bulletin

This security bulletin contains one low risk vulnerability.

1) Exposure of resource to wrong sphere

EUVDB-ID: #VU22282

Risk: Low

CVSSv3.1: 4 [CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L/E:U/RL:U/RC:C]

CVE-ID: CVE-2019-13546

CWE-ID: N/A

Exploit availability: No

Description

The vulnerability allows a local attacker to gain access to system resources.

The vulnerability exists due to the error that allows to bypass a lock screen of the Windows operating system when the software is running. An attacker with physical access to the system can bypass the Windows lock screen and gain unauthorized access to the computer.

Mitigation

Cybersecurity Help is currently unaware of any official solution to address this vulnerability.

Vulnerable software versions

IntelliSpace Perinatal: K

External links

http://ics-cert.us-cert.gov/advisories/icsma-19-297-01


Q & A

Can this vulnerability be exploited remotely?

No. The attacker should have physical access to the system in order to successfully exploit this vulnerability.

Is there known malware, which exploits this vulnerability?

No. We are not aware of malware exploiting this vulnerability.