Multiple vulnerabilities in Innokas Vital Signs Monitor VC150



Published: 2021-01-08
Risk Low
Patch available YES
Number of vulnerabilities 2
CVE ID CVE-2020-27262
CVE-2020-27260
CWE ID CWE-79
CWE-74
Exploitation vector Network
Public exploit N/A
Vulnerable software
Subscribe
Vital Signs Monitor VC150
Hardware solutions / Medical equipment

Vendor

Security Advisory

1) Stored cross-site scripting

Risk: Low

CVSSv3: 4.7 [CVSS:3.0/AV:N/AC:L/PR:L/UI:R/S:C/C:L/I:L/A:N/E:U/RL:O/RC:C] [PCI]

CVE-ID: CVE-2020-27262

CWE-ID: CWE-79 - Improper Neutralization of Input During Web Page Generation ('Cross-site Scripting')

Exploit availability: No

Description

The disclosed vulnerability allows a remote attacker to perform cross-site scripting (XSS) attacks.

The vulnerability exists due to insufficient sanitization of user-supplied data. A remote user can inject and execute arbitrary HTML and script code in user's browser in context of vulnerable website.

Successful exploitation of this vulnerability may allow a remote attacker to steal potentially sensitive information, change appearance of the web page, perform phishing and drive-by-download attacks.

Mitigation

Install updates from vendor's website.

Vulnerable software versions

Vital Signs Monitor VC150: before 1.7.15

CPE External links

https://ics-cert.us-cert.gov/advisories/icsma-21-007-01

Q & A

Can this vulnerability be exploited remotely?

Yes. This vulnerability can be exploited by a remote authenticated user via the Internet.

Is there known malware, which exploits this vulnerability?

No. We are not aware of malware exploiting this vulnerability.

2) Code injection

Risk: Low

CVSSv3: 4 [CVSS:3.0/AV:P/AC:L/PR:N/UI:N/S:U/C:N/I:H/A:N/E:U/RL:O/RC:C]

CVE-ID: CVE-2020-27260

CWE-ID: CWE-74 - Improper Neutralization of Special Elements in Output Used by a Downstream Component ('Injection')

Exploit availability: No

Description

The vulnerability allows an attacker to manipulate the device data.

The vulnerability exists due to insufficient filtration of user-supplied data when processing HL7 v2.x messages. A physically proximate attacker with a connected barcode reader can inject HL7 v2.x segments into specific HL7 v2.x messages via multiple expected parameters.

Mitigation

Install updates from vendor's website.

Vulnerable software versions

Vital Signs Monitor VC150: before 1.7.15

CPE External links

https://ics-cert.us-cert.gov/advisories/icsma-21-007-01

Q & A

Can this vulnerability be exploited remotely?

No. The attacker should have physical access to the system in order to successfully exploit this vulnerability.

Is there known malware, which exploits this vulnerability?

No. We are not aware of malware exploiting this vulnerability.