#VU66931 Improper access control in CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor - CVE-2022-36385

 

#VU66931 Improper access control in CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor - CVE-2022-36385

Published: September 2, 2022


Vulnerability identifier: #VU66931
Vulnerability risk: Low
CVSSv4.0: CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:U/U:Clear
CVE-ID: CVE-2022-36385
CWE-ID: CWE-284
Exploitation vector: Local access
Exploit availability: No public exploit available
Vulnerable software:
CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Software vendor:
Contec

Description

The vulnerability allows a local attacker to gain unauthorized access to otherwise restricted functionality.

The vulnerability exists due to improper access restrictions. An attacker with physical access can plug in a USB drive, perform a malicious firmware update and load arbitrary firmware on the target device.


Remediation

Cybersecurity Help is currently unaware of any official solution to address this vulnerability.

External links