Improper access control in CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor - CVE-2022-36385
Published: September 2, 2022
Vulnerability identifier: #VU66931
CSH Severity: Low
CVSS v4: 7 [CVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N]
CVE-ID: CVE-2022-36385
CWE-ID: CWE-284
Exploitation vector: Local access
Exploit availability:
No public exploit available
Vulnerability details
The vulnerability allows a local attacker to gain unauthorized access to otherwise restricted functionality.
The vulnerability exists due to improper access restrictions. An attacker with physical access can plug in a USB drive, perform a malicious firmware update and load arbitrary firmware on the target device.
Affected software
CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
How to mitigate CVE-2022-36385
Cybersecurity Help is currently unaware of any official solution to address this vulnerability.