Vulnerability identifier: #VU49511
Vulnerability risk: Low
CVSSv3.1: 6.6 [CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:L/A:L/E:U/RL:O/RC:C]
CVE-ID:
CWE-ID:
CWE-330
Exploitation vector: Local network
Exploit availability: No
Vulnerable software:
Dana Diabecare RS
Hardware solutions /
Medical equipment
AnyDana-i
Hardware solutions /
Medical equipment
AnyDana-A
Hardware solutions /
Medical equipment
Vendor: SOOIL Developments Co., Ltd
Description
The vulnerability allows a remote attacker to gain access to potentially sensitive information.
The vulnerability exists due to the communication protocol of the insulin pump and its mobile applications use deterministic keys. A remote attacker on the local network can brute-force the keys via Bluetooth Low Energy.
Mitigation
Install updates from vendor's website.
Vulnerable software versions
Dana Diabecare RS: All versions
AnyDana-i: All versions
AnyDana-A: All versions
External links
http://ics-cert.us-cert.gov/advisories/icsma-21-012-01
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the local network (LAN).
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.