#VU59167 Insufficiently protected credentials in Agilia Connect WiFi module of the pumps - CVE-2021-33843

 

#VU59167 Insufficiently protected credentials in Agilia Connect WiFi module of the pumps - CVE-2021-33843

Published: January 4, 2022


Vulnerability identifier: #VU59167
Vulnerability risk: Medium
CVSSv4.0: CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:N/VI:L/VA:N/SC:N/SI:N/SA:N/E:U/U:Green
CVE-ID: CVE-2021-33843
CWE-ID: CWE-522
Exploitation vector: Remote access
Exploit availability: No public exploit available
Vulnerable software:
Agilia Connect WiFi module of the pumps
Software vendor:
Fresenius Kabi

Description

The vulnerability allows a remote attacker to compromise the target system.

The vulnerability exists due to the default configuration page is accessible without authentication. A remote attacker can use this functionality to change the exposed configuration values such as network settings.


Remediation

Install updates from vendor's website.