#VU59167 Insufficiently protected credentials in Agilia Connect WiFi module of the pumps


Published: 2022-01-04

Vulnerability identifier: #VU59167

Vulnerability risk: Medium

CVSSv3.1: 4.6 [CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:L/A:N/E:U/RL:O/RC:C]

CVE-ID: CVE-2021-33843

CWE-ID: CWE-522

Exploitation vector: Network

Exploit availability: No

Vulnerable software:
Agilia Connect WiFi module of the pumps
Client/Desktop applications / Other client software

Vendor: Fresenius Kabi

Description

The vulnerability allows a remote attacker to compromise the target system.

The vulnerability exists due to the default configuration page is accessible without authentication. A remote attacker can use this functionality to change the exposed configuration values such as network settings.

Mitigation
Install updates from vendor's website.

Vulnerable software versions

Agilia Connect WiFi module of the pumps: D25


External links
http://ics-cert.us-cert.gov/advisories/icsma-21-355-01


Q & A

Can this vulnerability be exploited remotely?

Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the Internet.

Is there known malware, which exploits this vulnerability?

No. We are not aware of malware exploiting this vulnerability.