Vulnerability identifier: #VU66931
Vulnerability risk: Low
CVSSv3.1: 6.1 [CVSS:3.1/AV:P/AC:L/PR:N/UI:R/S:U/C:H/I:H/A:H/E:U/RL:U/RC:C]
CVE-ID:
CWE-ID:
CWE-284
Exploitation vector: Local
Exploit availability: No
Vulnerable software:
CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor
Hardware solutions /
Medical equipment
Vendor: Contec
Description
The vulnerability allows a local attacker to gain unauthorized access to otherwise restricted functionality.
The vulnerability exists due to improper access restrictions. An attacker with physical access can plug in a USB drive, perform a malicious firmware update and load arbitrary firmware on the target device.
Mitigation
Cybersecurity Help is currently unaware of any official solution to address this vulnerability.
Vulnerable software versions
CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor: All versions
External links
http://ics-cert.us-cert.gov/advisories/icsma-22-244-01
Can this vulnerability be exploited remotely?
No. The attacker should have physical access to the system in order to successfully exploit this vulnerability.
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.