Risk | High |
Patch available | NO |
Number of vulnerabilities | 3 |
CVE-ID | CVE-2025-0626 CVE-2025-0683 CVE-2024-12248 |
CWE-ID | CWE-912 CWE-200 CWE-787 |
Exploitation vector | Network |
Public exploit | Vulnerability #1 is being exploited in the wild. |
Vulnerable software |
CMS8000 Patient Monitor Hardware solutions / Medical equipment |
Vendor | Contec |
Security Bulletin
This security bulletin contains information about 3 vulnerabilities.
EUVDB-ID: #VU103464
Risk: High
CVSSv4.0: 8.7 [CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:A/U:Amber]
CVE-ID: CVE-2025-0626
CWE-ID:
CWE-912 - Hidden Functionality (Backdoor)
Exploit availability: No
DescriptionThe vulnerability allows a remote attacker to compromise vulnerable system
The vulnerability exists due to hidden functionality (backdoor) is present in software. A remote attacker can use this functionality to gain full access to the application and compromise the affected system.
Note, the vulnerability is being actively exploited in the wild.
Cybersecurity Help is currently unaware of any official solution to address this vulnerability.
Vulnerable software versionsCMS8000 Patient Monitor: All versions
CPE2.3 External linkshttps://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication
Q & A
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the Internet.
Is there known malware, which exploits this vulnerability?
EUVDB-ID: #VU103468
Risk: Medium
CVSSv4.0: 4.6 [CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:U/U:Green]
CVE-ID: CVE-2025-0683
CWE-ID:
CWE-200 - Exposure of sensitive information to an unauthorized actor
Exploit availability: No
DescriptionThe vulnerability allows a remote attacker to gain access to potentially sensitive information.
The vulnerability exists due to the affected product transmits plain-text patient data to a hard-coded public IP address when a patient is hooked up to the monitor. A remote attacker can gain unauthorized access to sensitive information on the system.
MitigationCybersecurity Help is currently unaware of any official solution to address this vulnerability.
Vulnerable software versionsCMS8000 Patient Monitor: All versions
CPE2.3 External linkshttps://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
Q & A
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the Internet.
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.
EUVDB-ID: #VU103469
Risk: High
CVSSv4.0: 8.1 [CVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:U/U:Amber]
CVE-ID: CVE-2024-12248
CWE-ID:
CWE-787 - Out-of-bounds write
Exploit availability: No
DescriptionThe vulnerability allows a remote attacker to compromise vulnerable system.
The vulnerability exists due to a boundary error when processing untrusted input. A remote attacker can send specially crafted UDP requests, trigger an out-of-bounds write and execute arbitrary code on the target system.
MitigationCybersecurity Help is currently unaware of any official solution to address this vulnerability.
Vulnerable software versionsCMS8000 Patient Monitor: CMS7.820.075.08/0.74(0.75) - smart3250-2.6.27-wlan2.1.7.cramfs
CPE2.3https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01
Q & A
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the Internet.
Is there known malware, which exploits this vulnerability?
No. We are not aware of malware exploiting this vulnerability.