Vulnerability identifier: #VU47912
Vulnerability risk: Medium
CVSSv4.0: 5.1 [CVSS:4.0/AV:N/AC:L/AT:N/PR:L/UI:N/VC:L/VI:H/VA:L/SC:N/SI:N/SA:N/E:U/U:Green]
CVE-ID:
CWE-ID:
CWE-22
Exploitation vector: Network
Exploit availability: No
Vulnerable software:
SpaceCom
Hardware solutions /
Other hardware appliances
Data module compact plus
Hardware solutions /
Other hardware appliances
Battery pack with Wi-Fi
Other software /
Other software solutions
Vendor: B. Braun Melsungen AG
Description
The vulnerability allows a remote attacker to perform directory traversal attacks.
The vulnerability exists due to input validation error when processing directory traversal sequences. A remote authenticated attacker can send a specially crafted HTTP request and upload arbitrary files.
Mitigation
Install update from vendor's website.
Vulnerable software versions
SpaceCom: U61 - L81
Data module compact plus: A10 - A11
Battery pack with Wi-Fi: U61 - L81
External links
https://ics-cert.us-cert.gov/advisories/icsma-20-296-02
https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/se...
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote authenticated user via the Internet.
Is there known malware, which exploits this vulnerability?