Vulnerability identifier: #VU48957
Vulnerability risk: Medium
CVSSv4.0: 5.9 [CVSS:4.0/AV:A/AC:L/AT:N/PR:N/UI:A/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:U/U:Green]
CVE-ID:
CWE-ID:
CWE-367
Exploitation vector: Local network
Exploit availability: No
Vulnerable software:
Smart Model 25000 Patient Reader
Mobile applications /
Apps for mobile phones
Vendor: Medtronic
Description
The vulnerability allows a remote attacker to execute arbitrary code on the target system.
The vulnerability exists due to a race condition in the MCL Smart Patient Reader software update system. A remote attacker on the local network can upload and execute unsigned firmware and execute arbitrary code on the system.
Mitigation
Install updates from vendor's website.
Vulnerable software versions
Smart Model 25000 Patient Reader: before 5.2
External links
https://ics-cert.us-cert.gov/advisories/icsma-20-345-01
Can this vulnerability be exploited remotely?
Yes. This vulnerability can be exploited by a remote non-authenticated attacker via the local network (LAN).
Is there known malware, which exploits this vulnerability?